BIOCOMPATIBILITY OF A PDMS-COATED MICRO-DEVICE: BLADDER VOLUME MONITORING SENSOR
PAPER|Updated:2023-01-03
|
BIOCOMPATIBILITY OF A PDMS-COATED MICRO-DEVICE: BLADDER VOLUME MONITORING SENSOR
BIOCOMPATIBILITY OF A PDMS-COATED MICRO-DEVICE: BLADDER VOLUME MONITORING SENSOR
Chinese Journal of Polymer Science2012年30卷第2期 页码:242-249
Affiliations:
Department of Urology, St. Vincent's Hospital, College of Medicine, the Catholic University of Korea, Suwon, Korea
Department of Urology, Seoul St. Mary's Hospital, College of Medicine, the Catholic University of Korea, Seoul, Korea
Department of Urology, Yeouido St. Mary's Hospital, College of Medicine, the Catholic University of Korea, Seoul, Korea
Department of Mechanical Engineering, Sogang University, Seoul, Korea
Author bio:
**Sae Woong Kim, E-mail: ksw1227@catholic.ac.kr
Funds:
Ministry of Education, Science and Technology(2009-0082114);Healthy Medical Treatment Research and Development Program of the Ministry of Health & Welfare(A090481)
Dong Sup Lee, Su Jin Kim, Jun Ho Sohn, 等. BIOCOMPATIBILITY OF A PDMS-COATED MICRO-DEVICE: BLADDER VOLUME MONITORING SENSOR[J]. Chinese Journal of Polymer Science, 2012,30(2):242-249.
Dong Sup Lee, Su Jin Kim, Jun Ho Sohn, et al. BIOCOMPATIBILITY OF A PDMS-COATED MICRO-DEVICE: BLADDER VOLUME MONITORING SENSOR[J]. Chinese Journal of Polymer Science, 2012, 30(2): 242-249.
Dong Sup Lee, Su Jin Kim, Jun Ho Sohn, 等. BIOCOMPATIBILITY OF A PDMS-COATED MICRO-DEVICE: BLADDER VOLUME MONITORING SENSOR[J]. Chinese Journal of Polymer Science, 2012,30(2):242-249.DOI: 10.1007/s10118-012-1119-1.
Dong Sup Lee, Su Jin Kim, Jun Ho Sohn, et al. BIOCOMPATIBILITY OF A PDMS-COATED MICRO-DEVICE: BLADDER VOLUME MONITORING SENSOR[J]. Chinese Journal of Polymer Science, 2012, 30(2): 242-249.DOI: 10.1007/s10118-012-1119-1.
BIOCOMPATIBILITY OF A PDMS-COATED MICRO-DEVICE: BLADDER VOLUME MONITORING SENSOR
We evaluated the biocompatibility of a dimethylpolysiloxane-coated micro-device which had been designed for monitoring real-time bladder volume in previous studies. The extract assay with dimethylpolysiloxane which had been used for coating the micro-device to measure the bladder volume was performed as an
in vitro
cytotoxicity test. For
in vivo
biocompatibility testing
the inflammatory responses around the implantation site of the micro-device in subcutaneous tissue of rat were assessed by light microscope with H
&
E stain and fluorescence microscope with ED1 stain and von Willebrand factor stain. The averages of cell viability in dimethylpolysiloxane group were 84.6% and 82.3% at 24 h and 72 h incubation
respectively. The qualitative evaluations with light and fluorescence microscope revealed that the inflammatory changes peaked during 2 weeks but almost disappeared at 4 weeks after implantation of devices. The quantitative evaluations for granulation layer formation and neovascularization showed that the thickness of the layer in dimethylpolysiloxane group peaked during 2 weeks but it came to be stabilized at 4 weeks as thin as at 2 weeks in control group
and the frequency of neovascularization was higher in dimethylpolysiloxane group than in control group but it was not increased with time. The dimethylpolysiloxane-coated micro-device is thought be a reliable bio-medical device.
关键词
Keywords
references
The trial reading is over, you can activate your VIP account to continue reading.